CONSTRUCTION
- Constructed and fabricated on fully automatic CNC machines.
- Chamber interior is designed with dull or mirror finish stainless steel.
- Fluorescent Tubes for daylight effect and UV tubes for ultraviolet ray in accordance with the ICH guidelines.
- External finish manufactured with Stainless Steel for GMP model and G.I powder coated for the standard model.
- The internals of the chamber is water and vapor tight.
- Seams are arc-welded for strength & durability.
- Chambers are regulated with air circulation, heating & cooling, & humidity generator system.
- Double Door system with outer door insulation. The inner door offers full internal visibility. Doors sealed with silicone gasket along with Lock & key feature.
- Easily accessible Control Panel with electrical & refrigeration switchgear mounted on top.
- Mounted on heavy castor wheels for easy movement.
- Adjustable Stainless Steel electro polished wire mesh type trays; Easy to clean.
OPTIONAL FEATURES
- Software (21 CFR PART-11 Compliance)
- Stand by Cooling System
- Stand by Humidity System
- Scanner (4 Temp. +4 RH for Online Monitoring Purpose)
- Mobile Alert System
- Security Alarm System
- Extra Tray
OPTION 1 (PID CONTROLLER BASED)
- Microprocessor based PID controller with auto tuning for accurate control of temperature and humidity conditions.
- Rotronic make 4-20mA sensors for better accuracy/ Pt100 Sensor for Temperature & Direct % RH electronic capacitor type humidity.
- Data Storage unit facility taao store data and to print directly on printer or transfer to software with Date, Time, SV, PV, Door Open/ Close, Power On/Off remarks.
- Safety: Temperature & Humidity High / Low Alarm, Temperature very High / Low cut off, Water level low cut off.
OPTION 2 (HMI + PLC CONTROLLER BASED)
- 4.3″ HMI with TFT LCD Colour touch screen with password protection for all parameter actual status monitoring of utilities like Sensor, Heating & Cooling System, Humidity System).
- Rotronic make 4-20mA sensors for better accuracy.
- PLC with auto tuning facility for accurate control of temperature and humidity conditions.
- PLC system for an auto changeover of standby systems.
- Data Storage unit facility to transfer the data to software via Ethernet for web based online remote monitoring with Date, Time, SV, PV, Door Open/ Close, Power On/Off remarks.
- Safety/Alert:
- Temperature & Humidity High / Low alarm, Temperature & Humidity High / Low cut off, Low water level cut off, Auto change over for standby systems.
- Software (21 CFR PART-11 Compliance) :
- Log data, Event data, Audit Trail data, E-Records & E-signatures, Graphical analysis& Data acquisition.
TECHNICAL SPECIFICATION
Specification Temperature Range 20 °C to 50 °C Accuracy + 1.0 °C Uniformity + 2.0 °C Control System PLC or PID Based SIZES
Models Capacity (in ltrs.) Dimensions (W X D X H) Nos. of Light Bank Internal (mm) Width Depth Height External (mm) Width Depth Height GMP-PSC227L 227 600 X 630 X 600 875 X 1065 X 1180 1 GMP-PSC400L 400 600 X 630 X 1060 875 X 1065 X 1640 2